FDA WARNING_LETTER - Applied Medical Resources Corp - December 16, 2014
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On April 10, 2015, the FDA issued a Warning Letter to Applied Medical Resources Corp following an inspection from November 3 to December 16, 2014. The inspection revealed the company's medical devices, used in various surgeries, were adulterated under 21 U.S.C. § 351(h) due to non-conformity with Quality System (QS) regulation (21 CFR Part 820).
Key violations include: 1. **Failure to implement CAPA procedures (21 CFR 820.100(a)(3)):** Inadequate analysis of quality data for Suction/Irrigator and Kii Trocar complaints, leading to unaddressed non-conformities and unverified/unvalidated CAPAs. 2. **Failure to investigate complaints (21 CFR 820.198(c)):** Inadequate root cause investigation for device failures, including patient deaths and serious injuries linked to Kii Shield Bladed System, Epix Universal Clip, and Suction/Irrigator. Test methods, raw data, and trending were insufficient. 3. **Failure in design verification (21 CFR 820.30(f)):** Design output test results for Kii Shielded Bladed Access System lacked raw data, rationale for safety factors, and justification for changes in acceptance criteria. 4.
ID · 9a1f0879-a13f-4a2f-91ac-0943f1aaf578