FDA WARNING_LETTER - Avacen Inc. - July 25, 2013
Discuss this record with AI
Document or profile
One record shows the event. A profile shows the pattern.
Buy this document for the source file and details behind this single event.
Choose full profile access to connect related records, sites, people, citations, and enforcement history in one view.
On October 23, 2013, the FDA issued a Warning Letter to Avacen, Inc. following an inspection from July 24-25, 2013. The inspection revealed that Avacen's product, The Avacen Thermal Exchange System (TATES), is an adulterated device under section 501(h) of the Act, as its manufacturing, packing, storage, or installation methods do not conform to the Quality System regulation (21 CFR Part 820).
Key violations include: - Failure to establish and maintain procedures for corrective and preventive action (21 CFR 820.100(a)). - Failure to establish and maintain adequate design control procedures (21 CFR 820.30(a)). - Failure to maintain complaint files and establish adequate complaint handling procedures (21 CFR 820.198(a)). The provided procedure dated 7/24/2013 was deemed inadequate. - Failure to establish and maintain procedures for acceptance activities (21 CFR 820.80(a)). - Failure to establish and maintain adequate purchasing control procedures (21 CFR 820.50). - Failure to establish and maintain procedures for Device History Records (DHRs) (21 CFR 820.184). - Failure to establish and maintain adequate procedures