WARNING LETTER
•Genesis Biosystems, Inc•June 1, 2015

FDA WARNING_LETTER - Genesis Biosystems, Inc - June 01, 2015

View on Dashboard

Your cart

Your cart is empty.

Browse public document pages and click Buy this document. We'll ask for your email at checkout, no account needed up front.

Browse documents

Discuss this record with AI

ChatGPT

Claude

Perplexity

Grok

Copilot

Record Details

An FDA inspection of Genesis Biosystems, Inc. from December 2014 to June 2015 revealed significant violations regarding their DermaFrac Micro-channeling System and Lipivage Fat Harvest and Transfer System. The DermaFrac System was found adulterated (21 U.S.C. § 351(f)(1)(B)) for lacking an approved PMA or IDE, and misbranded (21 U.S.C. § 352(o)) for failing to submit a 510(k) premarket notification. Its micro-channeling technology, differing from 510(k)-exempt dermabrasion devices, raised new safety questions.

Both devices were adulterated (21 U.S.C. § 351(h)) due to non-conformity with the Quality System regulation (21 CFR Part 820). Specific deficiencies included failure to validate computer software (21 CFR 820.70(i)) and inadequate Device History Record (DHR) procedures (21 CFR 820.184). The firm's January 2015 response was deemed inadequate, with corrective actions still ongoing.

Genesis Biosystems must immediately cease activities causing misbranding/adulteration of the DermaFrac System. A written response within fifteen business days is required, detailing corrective actions, prevention plans, and a timetable. Non-compliance risks regulatory actions like seizure, injunction, and civil money penalties. The letter also highlighted that Lipivage may require CBER premarket review, and DermaFrac's topical solution claims imply drug intended uses, requiring drug provision compliance.

Company
Genesis Biosystems, Inc
Inspection Date
June 1, 2015
Product Type
Devices
Office
Dallas District Office
Person
  • Reynaldo Rodriguez (District Director)
Open in Dashboard

ID · b4aa67d6-a951-4ade-8cac-14d55d6e6524

Violation Codes10
21 U.S.C. 321(h)21 U.S.C. 351(f)(1)(B)21 U.S.C. 360e(a)21 U.S.C. 360j(g)21 U.S.C. 352(o)21 U.S.C. 360(k)21 CFR 807.81(b)21 U.S.C. 351(h)21 CFR 820.70(i)21 CFR 820.184

Full citation text and observation details available on the Dashboard.

footer
Global Key Solutions
Regulatory Data Transformed
Contact
New York Office:
370 Jay Street 7th Floor
Brooklyn, NY 11201
Scranton Office:
Scranton Enterprise Center, Suite 217
201 Lackawanna Avenue
Scranton, PA, 18503
🇺🇸+1 917-789-5099
🇪🇺+48 532-447-507
info@globalkeysolutions.net
Quick Links
HomePlatform OverviewMedia
KeyPedia™ Lite
Browse HubDatabase DirectoryRecordsCompaniesOfficesPeopleCareersAbout UsFounding TeamMeet Our ExpertsPartners
Legal & Ethics
Privacy PolicyTerms of ServiceOur Values & MissionCode of Ethics & AI Policy
©2026, Global Key Solutions All rights reserved.